Emo‐3000 – AW Gear Meters EMO-3000 User Manual
Page 30

EMO‐3000
Operation and Programming Manual
29
3 = LINEARIZER
* * LINEARIZER PROGRAM * * * CH = XXX
POINT 01
‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 02
‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 03
‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 04
‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 05
‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 06
‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
F1 = ESC
F2 = PGUP
F3 = PGDN
F4 = PRGM
The table continues for 10 points and errors by F2 = PGUP.
POINT 07 ‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 08 ‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 09 ‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
POINT 10 ‐ ‐ ‐>
XXXX.X
ERROR = +XXX.XX
Programming the Linearizer is different from the other screens because it calls up the MENU BAR
(below) if you select F4 = PRGM.
F1 = ESC F2 = POINT F3 = ERROR
You must select either the point or the error percentage for programming. If you select F2 = POINT, the
MENU BAR shows the following choices:
F1 = ESC F2 = CLR F5 = DP F6 = ENT
Key in the correct deviation point and enter with F6. Correct any mistakes using F2 = CLEAR, F5 =
DECIMAL POINT.
If you select F3 = ERROR, the MENU BAR shows the following choices:
F1 = ESC F2 = CLR F3 = SGN F5 = DP F6 = ENT
Use “F3 = SGN” to key in the correct error percentage and +/‐ sign. Press F6 to enter.
4 = SPECIAL REPEAT
If you select 4 =SPECIAL, the DM‐3000 returns to the programming section last entered. This is a handy
feature if you need to program several channels in the same manner, or if you need to make quick
changes for variables commonly used.