4 retraction/filtration, 5 field enclosures, Retraction/filtration – LumaSense Technologies BoilerSpection-SD User Manual
Page 41: Field enclosures
BoilerSpection HD/SD Manual
Troubleshooting 41
5.4 Retraction/Filtration
Symptom
Probable Cause
Corrective Action
Immediate retraction
Loss of air, air pressure caused by site
supply failure
Check supply of air and power
Failure to retract
Mechanical failure of the retraction
system
Check mechanical fittings and
enclosure for loose or out of
place components and repair
Lens failure due to heat.
Melts within boiler
Replace or service lens
Damage to camera/lens
Poorly tuned retraction forces
damaged to the system
Service broken components or
properly tune retraction
system to avoid future damage
Loss of air pressure
Clogged filter caused filter failure
within the filtration unit
Replace Air filters
Failure of Vortex
Cooling
Premature failure of
filtering
Excessive heat to the filter system
caused by the environment
Consider adding shielding to
reduce the air filters
environmental exposure
Auto retracts, loss of
image
Site power failure
Verify and replace failed
component
Short Circuits, loss of
image
Water damage
Verify and replace failed
component
5.5 Field Enclosures
Symptom
Probable Cause
Corrective Action
Loss of image
Communication failures ie. Ethernet
failure, connections/switch failure
Open circuit breakers and
check all power wiring
connetions
Power failure to cabinet
Check integrity of input power
supply
Communication failure due to
mechanical vibrations
Check mechanical fittings and
enclosure for loose or out of
place components and repair
Cabinet left open causing
heat/water/dust in affect components
Verify and replace failed
component
Power surges
Site supply issues causing circuit
breakers to trip
Check integrity of input power
supply