BoilerSpection HD/SD Manual
Introduction 14
Each camera has its own independently operated retraction system. Included with each is a
regulator for the compressed air supply, pressure switch, check valve, ball valve, solenoid, and a
controls enclosure. The retract device also includes all air lines and connections that are required for
the system. It also includes a set of air filters for each camera/retract location to ensure that clean
air is supplied to the BoilerSpection lens and to the system's vortex coolers. Drawings can be found in
the Chapter 7 of this manual.
Figure 8: Retract Control Box
Figure 9: Retract Air Filtration Reservoirs
and Plumbing Assembly
2.3.3 Mounting Plate
Each camera includes a plate for mounting the retraction device to the furnace wall. Customers can
select between weld-on plates or bolt-on plates. Customers can also provide requirements to us for
custom plates to suit their furnace requirements.
2.3.4 Field Switch Enclosure (FSE)
The Field Switch Enclosure (FSE) provides a central
connection point for all communication cables from the
BoilerSpection cameras. The FSE includes an industrial
Ethernet switch with a fiber optic port to link to the control
room. The fiber optic link can be up to 250m (820') to the
Imaging Processor / Controller (IPC) server located in the
control room.
The customer is responsible for providing Ethernet
(e.g., Cat-5 or Cat-6) cables to connect the BoilerSpection
cameras to the FSE and also for routing and terminating the
fiber optic link to the control room.
Note: A circuit breaker is provided but not shown
in the photo.
2.3.5 Power Distribution Enclosure (PDE)
The Power Distribution Enclosure (PDE) is the central connection point for all BoilerSpection System
power supplies. This PDE is referred to as the “power panel” in the Technical Drawings, which are
found in Chapter 7.
The PDE is a configuration option and, depending on the system design at your facility, you may or
may not have the PDE from LumaSense. The PDE provides a central location for the connection of
AC power from the plant to power multiple cameras located on the same elevation or near one
Figure 10: Field Switch Enclosure (FSE)