BoilerSpection HD/SD Manual

Troubleshooting 40

Change in image color

Accuracy out of spec

Contact LumaSense for
calibration or evaluation

Grainy/poor

image

Noise

out

of

spec

Shifted Image

Alignment shifts within the enclosure
due to mechanical vibrations

Check mechanical fittings and
enclosure for loose or out of

place components and repair

Remote focus will not
adjust (HD systems only)

Mechanical failure – focusing motor
caused by improper tuning of
retraction system

Service camera

Loss of image

Loss of communication

Check connection to the
camera; investigate the
possibility of an IP address

corruption


5.3 Enclosure & Air-Purge

Symptom

Probable Cause

Corrective Action

Loss of cooling,
overheating, image

failure

Vortex coolers clogged or damaged
due to dirty air supply

Verify and replace failed
component

Thermostat failure within enclosure

Verify and replace failed

component

Solenoid failure

Verify and replace failed
component

Hose failure from cooler caused by
mechanical vibrations

Check all air connections

Hose failure from cooler caused by
environmental damage

Verify and replace failed hose

Overheating, internal

dust, image quality loss

Leakage in enclosure caused by

installation error

Check all gaskets and seals

Leakage in enclosure caused by
environmental damage

Replace gaskets and seals.
Investigate ways to mitigate
further environmental
exposure

Leakage in enclosure caused by

mechanical vibration

Check mechanical fittings and

enclosure for loose or out of
place components and repair