BoilerSpection HD/SD Manual
Troubleshooting 40
Change in image color
Accuracy out of spec
Contact LumaSense for
calibration or evaluation
Grainy/poor
image
Noise
out
of
spec
Shifted Image
Alignment shifts within the enclosure
due to mechanical vibrations
Check mechanical fittings and
enclosure for loose or out of
place components and repair
Remote focus will not
adjust (HD systems only)
Mechanical failure – focusing motor
caused by improper tuning of
retraction system
Service camera
Loss of image
Loss of communication
Check connection to the
camera; investigate the
possibility of an IP address
corruption
5.3 Enclosure & Air-Purge
Symptom
Probable Cause
Corrective Action
Loss of cooling,
overheating, image
failure
Vortex coolers clogged or damaged
due to dirty air supply
Verify and replace failed
component
Thermostat failure within enclosure
Verify and replace failed
component
Solenoid failure
Verify and replace failed
component
Hose failure from cooler caused by
mechanical vibrations
Check all air connections
Hose failure from cooler caused by
environmental damage
Verify and replace failed hose
Overheating, internal
dust, image quality loss
Leakage in enclosure caused by
installation error
Check all gaskets and seals
Leakage in enclosure caused by
environmental damage
Replace gaskets and seals.
Investigate ways to mitigate
further environmental
exposure
Leakage in enclosure caused by
mechanical vibration
Check mechanical fittings and
enclosure for loose or out of
place components and repair