7 orion manifold - panel installation, Orion manifold - panel installation – Condec DLR3110 User Manual

Page 47

background image

Maintenance and Service

43

6. Clean both the needle (11) and housing (10) in solvent, dry the parts and apply a small amount of

fluorinated Krytox grease to the needle threads before reassembly.

7. Assemble the Valve Needle (11) into the Valve Needle Housing (10) and turn it until it stops.

8. Reinstall the needle/housing assembly into the valve cavity until finger tight.

9. Mount the manifold body (16) in a vise. For the Coarse (pressure) and Vent valves only, torque the

needle/housing assembly to 325 in-lb. using the needle housing socket (PN 65580) and torque wrench.

10. Install the housing lock nuts (1) onto the housing (10) and tighten until snug with the 3/4" socket.

11. Using the .050" hex wrench, install and tighten the lock nut (2) and set screw (34).

12. Install the Knob Insert (4) over the Valve Needle (11) shaft, align the Set Screws (23) with the indents

and tighten with the .061" hex wrench.

13. For the isolation valves (two inner valves), install the Needle Housing (19) and tighten until snug using

the Isolation Valve Housing Installation Socket (PN 59793), Female Socket (PN 65581) and torque
wrench.

NOTE: There is no specified torque, so use care when tightening so as not to break the socket nibs.

14. Install the Gear (6) over the Isolation Valve Needle (18) shaft, align the Set Screws (23) with the indents

and tighten with the .061" hex wrench.

15. Apply a small amount of fluorinated Krytox grease to the threads of the isolation valve needles (18) and

install into the valve by turning counter-clockwise. Rotate the gear until the needle just stops at the seat.

7.2.7

ORION Manifold - Panel Installation

Tools required:

7/16" open end wrench
Phillips screwdriver
hex wrench (.061")
SNOOP, liquid leak gas detector (PN 64781)

11/32" open end wrench (thin)

1. If not already done, remove the panel knobs from the Input (pressure), Vernier and Vent valves using the

.061" hex wrench.

2. Install the manifold with the transducer port side facing the panel bottom. Install the two mounting

screws (PN 60837) from the panel front and tighten until snug.

3. Install the Vernier Knob (17) onto the Vernier Valve Shaft (14). Align the set screws (25) with the

indentations on the Vernier Valve Shaft and tighten until snug using the .061" hex wrench.

4. To install and adjust the Input (pressure) and Vent Valve Knobs, follow the procedure in Section 7.2.8.

5. Install the transducer into the manifold port, tighten with the 11/32" thin wrench and reconnect its wire

connector.

NOTE: If transducer is hard-wired, connect the 4 wires to the terminal block, TB1, on the CPU board per the
following:

6. Install all tubing sections that attach to the ORION Manifold.

7. Install cover as described in Section 7.2.1 on page 39.

Transducer Wires

Terminal Block Wires

+ Excitation

TB1-1 Green wire

- Excitation

TB1-6 Black wire

+ Signal

TB1-3 White wire

- Signal

TB1-4 Red wire

Table 7-2. Transducer Wiring Specification