The alpha-numeric display reads, Key on the display panel, Save the calibration data by opening rocker switch – Condec DLR3110 User Manual
Page 15: Position, Hyst. cal target=xxx, Enter, Filter 5, Figure 4-1. instrument calibration set-up, Re-calibration only 11
Re-Calibration Only
11
9. The alpha-numeric display reads
HYST. CAL TARGET=XXX
(X represents the preset target value on
display). Set the input pressure from the test standard to the target value on the display. Wait for system
to stabilize and press the
ENTER
key on the display panel.
NOTE: For ABS models, use a vacuum pump to reach the applicable SPAN or HYST target values that are
below local barometric pressure.
10. The alpha-numeric display reads
FILTER 5
. Save the calibration data by opening rocker switch
S1
position
2
. DLR3110 returns to normal mode.
11. Vent system and remove test standard.
Figure 4-1. Instrument Calibration Set-Up
NOTE: AC input and input port are on back side rack-mountable calibrator.