Configuration and Calibration
15
5.5.2
Auto Set-Up Menu (Gage)
See Section 9.0 on page 63 for a complete listing of the default settings for all gage and absolute selections under
the AUTO set-up mode.
NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit
menu (see Table 5-3 on page 14).
LED Display
Alpha-Numeric Display
Key Pressed
Remarks
SET
AUTO ?
YES key
----
GAGE ?
YES key
SET-UP
3 PSIG ?
NO key
Continue to press NO key until the
alpha-numeric display shows the
DLR3110’s full scale range.
YES key
Press YES key when the alpha-numeric
display shows the DLR3110’s full scale
range.
SET-UP
ACCEPT ?
YES or NO.
NO key
Continue to press NO key until the
alpha-numeric display shows the
DLR3110’s full scale range. See note
below.
YES key
Accept data
XXXX
CAPACITY
YES key
“XXX” shown on LED display represents
the full scale range and is based on
previous selection.
—
EXIT ?
YES key
Return to SET-UP/ I/O?. (see Section 5.4
on page 13).
NO key
Returns to XXX/CAPACITY
Table 5-4. Auto Set-Up Mode (Gage) Parameters Selections