6

DLR3110 Rack Mount Digital Pressure Calibrator Manual

2.6.1

Freeze Mode Specifications

The following bullet points outline general specifications for the DLR3110’s Freeze mode.

The freeze input “freezes”

the display for approximately five seconds on the rise and fall time of the

input (edge not level triggered) and can not be re-triggered.

Input pulse > 100 ms required to guarantee detection.

Freeze mode “freezes”

the 6 digit numeric and LED bar displays in whatever mode it is in (for example,

Normal, Net, Min and Max mode).

Alphanumeric display prompts

FREEZE

.

During the “freeze”

the front panel keys are inactive.

Inputs are a +5 VDC logic level. Sinking current approximately

0.5 mA.

2.7

Min and Max Mode

The Min and Max mode is designed to capture and store the highest and the lowest pressure readings. These
values may be recalled at anytime in the normal operating mode by repeatedly pressing the

DISPLAY

key until

the desired value is displayed. When either the minimum or maximum value is selected, the alphanumeric
display flashes between the mode (Min and Max) and its units.

NOTE: The Min and Max mode pressure data is automatically converted to any of the active conversion units by
stepping through the conversion units using the UP or DOWN arrow keys.

Both the Min and Max mode are independent of each other and are enabled (turned ON) in the Configuration
mode (see Section 5.11 on page 21).

2.7.1

Resetting Min and Max Values

To reset the stored minimum or maximum value:

1. Select the Min or Max display to be reset using the

DISPLAY

key.

2. Press the

CE

key. One of the following messages appears on the alphanumeric

display:

RESET MAX?

YES OR NO; RESET MIN? YES OR NO.

3. At the display prompt shown above, press the

YES

key to reset the value or press the

NO

key to abort

resetting the value.

NOTE: Min and Max data work with the standard pressure, not delta pressure data.

2.8

Remote Display Mode

The following describes the setup for connecting the optional six-digit remote display (PN 74683). The display
duplicates the DLR3110 six-digit front panel display for all ranges and units selected (see Figure 2-5 on page 7).

NOTE: Remote display does not show current unit or mode designation.

See Table 5-10 on page 24 for setting the following parameters:

Parameter

Configuration

Port 1

SIMPLEX

Baud

9600

Parity

EVEN

Port 1

Modify

delay 1

AUTO

t –d 1

OFF

data 1

COMPRESS

EOL 1

CR

SOt 1

NONE

EOt 1

NONE

Table 2-1. DLR3110 Parameters