DLR3110 Rack Mount Digital Pressure Calibrator Manual

Figure 5-3. Auto Set-Up Mode

5.5

Auto Set-Up Configuration Mode Selection

The auto set-up automatically sets the following conditions in the DLR3110:

SET-UP - CONFIG

configuration parameters

SET-UP - CONV

conversion parameters

SET-UP - I/O

configuration of inputs/outputs

---- CAL

sets capacity, count by, abs/gage, and cal units; resets all calibration values; sets all cal target

values for span and hysteresis.

5.5.1

Auto Set-Up Menu (General)

* Press YES key to accept selection; NO key to reject.

Upon completion and accepting an AUTO set-up configuration, the program automatically advances the
indicator to the calibration mode (see Section 5.6.1 on page 17).

LED Display

Alpha-Numeric Display

Key Pressed*

Remarks

SET

AUTO ?

YES key

See Section 5.4 on page 13

----

GAGE ?

YES key

See Section 5.5.2 on page 15

NO key

Continue to ---/ABS?

----

ABS ?

YES key

See Section 5.5.3 on page 16

NO key

Continue to ---/SPECIAL?

----

SPECIAL. ?

NO key

Allocated for future use, continue to
----/CUSTOM ?

----

CUSTOM ?

NO key

Allocated for future use, continue to
----/EXIT ?

----

EXIT ?

YES key

Return to main set-up mode selections
(see Section 5.4 on page 13).

NO key

Returns to ----/GAGE ?

Table 5-3. Auto Set-Up Mode Selection Parameters