3 auto set-up menu (absolute), 6 calibration set-up configuration mode selection, Auto set-up menu (absolute) – Condec DLR3110 User Manual
Page 20
16
DLR3110 Rack Mount Digital Pressure Calibrator Manual
5.5.3
Auto Set-Up Menu (Absolute)
See Section 9.0 on page 63 for a complete listing of the default settings for all gage and absolute selections under
the AUTO set-up mode.
NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit
menu (see Table 5-3 on page 14).
5.6
Calibration Set-Up Configuration Mode Selection
Use one of the following methods for configuring the calibration set-up mode:
Method 1
CAL (SELECT). See Section 5.4 on page 13 for information on selecting main mode settings.
Method 2
CAL (DIRECT). Closing only rocker switch position
2
on switch
S1
places the DLR3110 directly into
CALIBRATION mode. See Section 5.3 on page 12.
NOTE: If the AUTO SET-UP selection is accepted and completed, the indicator automatically advances to the calibration
mode (see Section 5.5 on page 14).
LED Display
Alpha-Numeric Display
Key Pressed
Remarks
SET
AUTO ?
YES key
----
GAGE ?
NO key
—
ABS ?
YES key
SET-UP
15 PSIA
NO key
Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range.
YES key
Press YES key when the alpha-numeric
display shows the DLR3110’s full scale
range.
SET-UP
ACCEPT ?
YES or NO
NO key
Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range. See note
below.
—
—
YES key
Accept data
XXXX
CAPACITY
YES key
“XXX” shown on LED display represents
the full scale range and is based on
previous selection.
—
EXIT ?
YES key
Return to SET-UP/ I/O?. (see Section 5.4
on page 13).
NO key
Return to XXX/CAPACITY
Table 5-5. Auto Set-Up Mode (Absolute) Parameters Selections