DLR3110 Rack Mount Digital Pressure Calibrator Manual

5.5.3

Auto Set-Up Menu (Absolute)

See Section 9.0 on page 63 for a complete listing of the default settings for all gage and absolute selections under
the AUTO set-up mode.

NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit
menu (see Table 5-3 on page 14).

5.6

Calibration Set-Up Configuration Mode Selection

Use one of the following methods for configuring the calibration set-up mode:

Method 1

CAL (SELECT). See Section 5.4 on page 13 for information on selecting main mode settings.

Method 2

CAL (DIRECT). Closing only rocker switch position

2

on switch

S1

places the DLR3110 directly into

CALIBRATION mode. See Section 5.3 on page 12.

NOTE: If the AUTO SET-UP selection is accepted and completed, the indicator automatically advances to the calibration
mode (see Section 5.5 on page 14).

LED Display

Alpha-Numeric Display

Key Pressed

Remarks

SET

AUTO ?

YES key

See Section 5.5 on page 14

----

GAGE ?

NO key

See Section 5.5 on page 14

ABS ?

YES key

See Section 5.5 on page 14

SET-UP

15 PSIA

NO key

Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range.

YES key

Press YES key when the alpha-numeric
display shows the DLR3110’s full scale
range.

SET-UP

ACCEPT ?

YES or NO

NO key

Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range. See note
below.

YES key

Accept data

XXXX

CAPACITY

YES key

“XXX” shown on LED display represents
the full scale range and is based on
previous selection.

EXIT ?

YES key

Return to SET-UP/ I/O?. (see Section 5.4
on page 13).

NO key

Return to XXX/CAPACITY

Table 5-5. Auto Set-Up Mode (Absolute) Parameters Selections