3.1 Error Messages

When an error message appears, take appropriate measures by referring to the table
below.

Error Messages

Message

Corrective Action

Reference

Section

User name or Password is not right or

recorder is already connected by other

program!

In the communication settings, check that the user name and

password match with the settings on the instrument to be connected.

In addition, check that other software programs are not performing

communications with the instrument to be connected.

2.2

Connection error, please check

communication setting!

In the communication settings, check that the communication mode

and parameters match with those of the instrument to be connected.

2.2

Connection timeout.

There may be too much traffic.
Retry after a little while.

Failed to open file.

If the file cannot be loaded the second time, the file may be corrupt.

Select another file.

2.3

Failed to send data.

Displayed when the transmission of the setup data fails.
Check that the system configuration matches that of the connected

instrument.

2.9

Failed to make file.

Check the capacity of the directory, or check that other programs are

not using it.

The setting information is mismatch

current setting. Please select again.

A unsupported file is selected. Check the file.
The extension to setup data files is .pul.

Now recording. Can’t send settings.

Stop the recording on the µR before sending the data.

Now calculating. Can’t send settings.

Stop the computation on the µR before sending the data.

Now recording & calculating. Can’t send

settings.

Stop the recording and computation on the µR before sending the

data.

Sending data is not permitted to the

current user level.

Change the user making communication settings to administrator

level.

2.2

There is no format string.

Specify the character string in the message format.

2.6

Format string exceeds limit.

The character string written in message format is too long.

2.6

Printed string exceeds limit.

The character string written in message format is too long

when it is actually printed.

2.6

Format string is wrong.

There is an error in the message format string.

2.6

Channel selection is wrong.

There is an error in the channel designation in the message format.

2.6

Message selection is wrong.

There is an error in the message designation in the message format. 2.6

Warnings

Message

Some data has been modified, continue sending data?
System configuration has been changed. The input configuration and data will be initialized. Continue?
Contains invalid data. Open this setting?
Hardware and software configurations don’t match. Continue sending data?
Any destroyed A/D converter exists. Any settings may be failed to store.
This recorder doesn’t match all, Continue or not?
File version x.xx is not supported. Setting data will be changed to latest version x.xx. Do you continue operation?
Version x.xx is not supported. Setting data will be received as latest version x.xx. Do you continue operation?
This device is not supported. Setting data will be received as supported version x.xx device. Do you continue operation?
Connected device is not supported. Some setting data may not be send. Do you continue sending?
Connected device is not supported hardware.

Chapter 3