9 sending setup data to the recorder, 9 sending setup data to the recorder -36 – Yokogawa µR20000 User Manual
Page 49

2-36
IM 04P01B01-61E
2.9 Sending Setup Data to the Recorder
1.
From the
Communication menu, choose Send Setting. You can also click the
Send setting icon on the toolbar. The
Sending setting data dialog box opens.
Send setting icon
2.
Click
OK.
The transmission starts. When the transmission of the settings is complete, a
message appears to indicate it.
3.
Click
OK.
Note
• If a message appears, see section 3.1, “Error Messages.”
• If the difference between values of the correction points, or the difference between values
after correction, are small, for example several digits, an error may occur when you send the
setup data to the recorder. In that case, change the set values.