Anritsu 682XXB User Manual
Page 470
A-232
682XXB/683XXB MM
TEST
MODEL
RECORD
68287B/68387B
3-11 Power Level Accuracy and Flatness Tests
(Model 68287B/68387B without Option 2D Step Attenuator)
Power Level Accuracy Test Procedure
Set F1 to 5.0 GHz
Set F1 to 25.0 GHz
Set F1 to 55.0 GHz
Set L1 to:
Measured Power *
Set L1 to:
Measured Power *
Set L1 to:
Measured Power *
+10 dBm
+ 9 dBm
+ 8 dBm
+ 7 dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
+ 2 dBm
+ 1 dBm
0 dBm
– 1 dBm
– 2 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
– 9.5 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+ 2 dBm
+ 1 dBm
+ 0 dBm
– 1 dBm
– 2 dBm
– 3 dBm
– 4 dBm
– 5 dBm
– 6 dBm
– 7 dBm
– 8 dBm
– 9 dBm
– 10 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
* Specification is
±
1.0 dB.
* Specification is
±
1.0 dB.
* Specification is
±
1.5 dB.
Power Level Flatness Test Procedure (Step Sweep)
Set L1 to:
Max Power
Min Power
Variation **
+ 2 dBm
_________dBm
__________dBm
_________dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 60 GHz).
Power Level Flatness Test Procedure (Analog Sweep) (Model 68387B only)
Set L1 to:
Max Power
Min Power
Variation ***
+ 2 dBm
_________dBm
__________dBm
_________dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 40 GHz); 5.0 dB (40 to
60 GHz)(typical, not a specification).