3-14 Pulse Modulation Tests: Rise Time, Fall Time, Overshoot, and Level (Continued)
Pulse Level Accuracy (5 GHz, Pulse Width = 0.5
m
s)
Lower Limit
Measured Value
Upper Limit
Set F1 to 5.0 GHz
Set pulse width W1 to 5.0
m
s
Record the displayed output level L1 (step 5.b) . . . . . . . . . . . . . . . . . . . .
Record the displayed output level L1 (step 5.d) . . . . . . . . . . . . . . . . . . . .
_______ dBm
_______ dBm
Calculate and record the pulse level error. . . . . . . . . . . . . .
–1.0 dB
_______ dB
+1.0 dB
Pulse Level Accuracy (5 GHz, Pulse Width = 5.0
m
s)
Set F1 to 5.0 GHz
Set pulse width W1 to 5.0
m
s
Record the displayed output level L1 (step 5.b) . . . . . . . . . . . . . . . . . . . .
Record the displayed output level L1 (step 5.d) . . . . . . . . . . . . . . . . . . . .
_______ dBm
_______ dBm
Calculate and record the pulse level error. . . . . . . . . . . . . .
–0.5 dB
_______ dB
+0.5 dB
3-15 Pulse Modulation Test: Video Feedthrough
Test Procedure
Lower Limit
Measured Value
Upper Limit
Set F1 to 5.0 GHz
Set pulse width W1 to 5.0
m
s
Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes . . . .
–10 mV
_______ mV
+10 mV
TEST
MODEL
RECORD
68265B/68365B