682XXB/683XXB MM
A-119
TEST
MODEL
RECORD
68263B/68363B
3-11 Power Level Accuracy and Flatness Tests
(Model 68263B/68363B without Option 2B Step Attenuator)
Power Level Accuracy Test Procedure
Set F1 to 5.0 GHz
Set F1 to 25.0 GHz
Set L1 to:
Measured Power *
Set L1 to:
Measured Power *
+ 9 dBm
+ 8 dBm
+ 7 dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
+ 2 dBm
+ 1 dBm
+ 0 dBm
– 1 dBm
– 2 dBm
– 3 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
+ 2dBm
+ 1 dBm
+ 0 dBm
– 1 dBm
– 2 dBm
– 3 dBm
– 4 dBm
– 5 dBm
– 6 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
* Specification is
±
1.0 dB.
* Specification is
±
1.0 dB.
Power Level Flatness Test Procedure (Step Sweep)
Set L1 to:
Max Power
Min Power
Variation **
+ 6 dBm
_________dBm
__________dBm
_________dB
** Maximum variation is 1.6 dB.
Power Level Flatness Test Procedure (Analog Sweep) (Model 68363B only)
Set L1 to:
Max Power
Min Power
Variation ***
+ 6 dBm
_________dBm
__________dBm
_________dB
*** Maximum variation is 2.0 dB (2 to 20 GHz); 4.0 dB (20 to 40 GHz)(typical, not a specification).