Performance phase verification modulation tests – Anritsu 682XXB User Manual
Page 109
WIDE
FM Mode
1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing
SYSTEM
,
then
Reset
. Upon reset, the CW Menu is dis-
played.
b. Press
Edit F1
to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
2. Set up the Spectrum Analyzer as follows:
a. CF: 5.0 GHz
b. Span/Div: 100 kHz (1 MHz across the display)
3. On the 682XXB/683XXB, make the following set-
tings:
a. Press
MODULATION
, then
FM . At the re-
sulting
FM Status menu, press
More
to go the
additional
FM Status menu.
b. At the additional
FM Status menu, press
Wide
, then press Previous Menu to return to
the main
FM Status menu.
c. At the
FM Status menu, press
Edit Dev.
and
set the deviation for 2.40 rad.
d. Press
Edit Rate
and set the rate to 99.8 kHz.
e. Press
On/Off
to turn
FM on.
4. On the 682XXB/683XXB, use the cursor control
key to place the cursor under the last 0 in the de-
viation setting (2.40 rad). While observing the
first Bessel null (Figure 3-15), increase/decrease
the
FM deviation with the cursor control key to
reduce the carrier as low as possible.
5. Record the
FM deviation, displayed on the
682XXB/683XXB, on the Test Record.
6. Compute the
FM accuracy with the following for-
mula:
Accuracy in
M Deviation in rad
(
%)
(
)
.
=
ж
и
з
ц
ш
ч ´
F
240
100
7. The calculated result should be between 93.7%
and 106.3%. Record this result on the Test Re-
cord.
682XXB/683XXB MM
3-49
PERFORMANCE
PHASE
VERIFICATION
MODULATION TESTS
CENTER FREQUENCY
BESSEL
NULL
Figure 3-15.
Typical Spectrum Analyzer Display
of Bessel Null on
.M Waveform
NOTE
You may need to adjust the RBW
setting on the Spectrum Analyzer in
order to see the >–48 dBc null.