Alc slope calibration calibration – Anritsu 682XXB User Manual

Page 135

background image

3. Make the following selections on the 562 Network

Analyzer to normalize the step sweep.

a. Press CALIBRATION and follow the menu on

the display.

b. Press AUTOSCALE.

c. Press OFFSET/RESOLUTION and set the

Resolution to 0.2 dB.

4. On the 683XXB, press

Analog

to select the ana-

log sweep mode.

5. Adjust the ALC Slope DAC for frequencies

£2 GHz as follows:
a. At the

$

prompt on the PC display, type:

calset

and press .

b. Type:

1

and press .

c. Select item 10 (ALC Slope Het) from the menu.

Follow the instructions presented at the bot-
tom of the screen to enter a new DAC setting
value.

d. Enter different DAC setting values to find the

setting that adjusts the slope so that the power
at the start and stop frequencies match the
normalized straight line in step sweep mode.

e. When finished, exit the program.

f. Type:

X

and press to exit the calset

menu. (The

$

prompt will appear on the

screen.)

g. Record step completion on the Test Record.

6. Set up the 683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

then

Reset

. Upon reset the CW Menu is dis-

played.

b. Press

Edit L1

and set L1 to 5 dB less than

maximum leveled power for the instrument be-
ing calibrated (refer to Table 3-2, page 3-6).

c. Press

Step

. The Step Sweep Menu is dis-

played.

d. Press

Edit F1

and set the F1 start frequency

to 2.0 GHz.

e. Press

Edit F2

and set the F2 stop frequency

to upper limit of the 683XXB.

682XXB/683XXB MM

4-25

ALC SLOPE

CALIBRATION

CALIBRATION