Alc slope calibration calibration – Anritsu 682XXB User Manual
Page 139
d. Press
CW/SWEEP SELECT
to return to the
Step Sweep Menu display.
e. Press
Sweep Ramp
. At the resulting Step
Sweep Ramp Menu, press
Num of Steps
and
set the number of steps to 400.
3. Make the following selections on the 562 Network
Analyzer to normalize the step sweep.
a. Press
CALIBRATION
and follow the menu on
the display.
b. Press
AUTOSCALE
.
c. Press
OFFSET/RESOLUTION
and set the Reso-
lution to 0.5 dB.
4. On the 683XXB, press
Analog
to select the ana-
log sweep mode.
5. Adjust the ALC Slope as follows:
a. At the
$
prompt on the PC display, type:
slpcal
and press
On the computer keyboard, the adjustment
keys are:
Slope
E (Up)
D (Down)
(all bands)
Offset
Q (Up)
A (Down)
(band 1-4 only)
682XXB/683XXB MM
4-29
ALC SLOPE
CALIBRATION
CALIBRATION
8.4 GHz
20 GHz
40 GHz
ALC Slope Before Adjustment
Normalized
Step Sweep
50. 60, or
65 (67) GHz
2 GHz
Figure 4-8.
ALC Slope Adjustment Waveform Display