beautypg.com

Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 197

background image

682XXB/683XXB MM

5-31

Step 6.

Using the spectrum analyzer, verify the presence of the
+3 dBm

±5 dB, 10 MHz signal at the end of the cable.

G

If present, replace the A4 PCB.

G

If not present, go to step 7.

Step 7.

Disconnect cable W105 at A3J4.

Step 8.

Using the spectrum analyzer, verify the presence of the
+3 dBm

±5 dB, 10 MHz signal at A3J4.

G

If present, replace the cable W105.

G

If not present, replace the A3 PCB.

Step 9.

Reconnect cable W121 to A5J5 and disconnect cable W107 at
A5J1.

Step 10. Set up the 682XXB/683XXB to generate the CW frequencies

listed in Table 5-10.

Step 11. Using a spectrum analyzer, measure the frequency and am-

plitude of the signal at A5J1 for each of the CW frequencies
generated. In each case, the signal amplitude should be
+3 dBm

±3 dB with sidebands at <–65 dBc.

G

If the signals are correct in both frequency and amplitude,
go to step 12.

G

If the signals are incorrect, replace the A5 PCB.

Step 12. Reconnect cable W107 to A5J1 and run self-test again.

G

If error 111 is not displayed, the problem is cleared.

G

If error 111 is still displayed, contact your local

ANRITSU

service center for assistance.

Table 5-9.

Error Message 111 (2 of 2)

TROUBLESHOOTING

TROUBLESHOOTING

TABLES

682XXB/683XXB

CW Frequency

Measured Frequency

at A5J1

10.102 GHz

22 MHz

10.110 GHz

30 MHz

10.120 GHz

40 MHz

Table 5-10.

Fine Loop Frequencies