Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual
Page 217
682XXB/683XXB MM
5-51
Error 144 RF was Off when Selftest started. Some tests where
not performed
Description: Indicates that some self-tests where not performed be-
cause the RF Output was selected
OFF
on the front panel.
Step 1.
Press the
OUTPUT
key on the front panel to turn the RF
Output
ON
.
Step 2.
Run self-test again.
Table 5-24.
Error Message 144
TROUBLESHOOTING
TROUBLESHOOTING
TABLES
Error 145 AM Meter or associated circuitry failed
Description: Indicates a failure of the internal amplitude modulation
function. The 682XXB/683XXB may or may not provide amplitude
modulation of the RF output signal using modulating signals from an
external source.
Step 1.
Set up the 682XXB/683XXB as follows:
a.
Press
SYSTEM
, then
Reset
.
b.
Press
MODULATION
, then
AM
to go to the Internal AM
Status display.
c.
Press
On/Off
to turn internal amplitude modulation on.
Step 2.
Using an oscilloscope, verify the presence of a 10 volt peak to
peak sine wave signal with a period of 100
ms at the rear
panel
AM OUT
connector.
G
If present, replace the A10 PCB.
G
If not present, replace the A8 PCB.
Table 5-25.
Error Message 145