CALIBRATION
FOLLOWING
SUBASSEMBLY
REPLACEMENT
Table 4-2 (page 4-6) lists the calibration that should be performed fol-
lowing the replacement of 682XXB/683XXB subassemblies or RF com-
ponents.
4-5
CONNECTOR AND KEY
LABEL NOTATION
The calibration procedures include many references to equipment in-
terconnections and control settings. For all 682XXB/683XXB refer-
ences, specific labels are used to denote the appropriate menu key,
data entry key, data entry control, or connector (such as
CW/SWEEP
SELECT
or
RF OUTPUT
). Most references to supporting test equip-
ment use general labels for commonly used controls and connections
(such as Span or RF Input). In some cases, a specific label is used that
is a particular feature of the test equipment listed in Table 4-1.
4-4
682XXB/683XXB MM
CALIBRATION FOLLOWING
CALIBRATION
SUBASSEMBLY REPLACEMENT
INSTRUMENT
CRITICAL
SPECIFICATION
RECOMMENDED
MANUFACTURER/MODEL
PROCEDURE
NUMBER
Frequency Counter
Frequency Range: 1 to 20 GHz
Input Impedance: 50
W
Resolution: 1 Hz
EIP Microwave, Inc. Model 578B
4-7
Spectrum Analyzer
Frequency Range: 1 to 20 GHz
Resolution Bandwidth: 10 Hz
Tektronix, Model 2794
4-13, 4-14
Power Meter,
with
Power Sensor
Power Range:–30 to +20 dBm
(1
m
W to 100mW)
Hewlett-Packard Model 437B, with
Power Sensor:
HP 8487A (0.01 to 50 GHz)
4-12
Power Meter
with
Power Sensor
Power Range:–30 to +20 dBm
(1
m
W to 100mW)
Anritsu ML4803A, with
Power Sensor:
MP716A4 (50 to 75 GHz
4-12
Frequency Standard
Frequency: 10 MHz
Accuracy: 1
´
10
–10
parts/day
Spectracom Corp., Model 8161
4-7
Function Generator
Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms
Hewlett-Packard, Model 8116A
4-12, 4-13,
4-14
Function Generator
Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms
Hewlett-Packard, Model 33120A
4-12. 4-13.
4-14
Digital Multimeter
Resolution: 4-1/2 digits (to 20V)
DC Accuracy: 0.002% +2 counts
DC Input Impedance: 10 M
W
AC Accuracy: 0.07% +100 counts
(to 20 kHz)
AC Input Impedance: 1 M
W
John Fluke, Inc.,Model 8840A, with
Option 8840A-09K (True RMS AC)
4-12, 4-13,
4-14
Table 4-1.
Recommended Test Equipment for Calibration Procedures (1 of 2)