Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual
Page 206
5-40
682XXB/683XXB MM
G
If the RF signal is correct in both frequency and ampli-
tude throughout the full sweep, go to step 8.
G
If there is no RF signal for all or part of the sweep or if
the amplitude of the RF signal is low, go to step 5.
Step 5.
Connect the X input of an oscilloscope to the 682XXB/
683XXB rear panel
HORIZ OUT
connector.
Step 6.
Using the oscilloscope, check for a –0.2 to –3.5 volt YIG
tuning ramp at A13TP10.
G
If the ramp signal is correct, go to step 7.
G
If the ramp signal is incorrect or not present, replace the
A13 PCB.
Step 7.
Using the oscilloscope, check for the YIG bias voltages at the
test points shown in Table 5-19.
G
If the YIG bias voltages are correct, replace the YIG-
tuned oscillator assembly.
G
If the YIG bias voltages are incorrect, replace the A13
PCB.
Step 8.
Run self-test again.
G
If no error message is displayed, the problem is cleared.
G
If any of the error messages, listed above, are displayed,
contact your local
ANRITSU
service center for assistance.
Table 5-18.
Error Messages 124, 125, and 126 (2 of 2)
TROUBLESHOOTING
TROUBLESHOOTING
TABLES
Test Point
YIG-tuned Oscillator Bias Voltages
2 to 8.4 GHz
8.4 to 20 GHz
A13TP3
+6V
+6V
A13TP5
0V
+8V
A13TP6
–5V
–5V
A13TP7
+8V
0V
Table 5-19.
YIG-tuned Oscillator Bias Voltages