background image

4. Connect the 682XXB/683XXB RF OUTPUT to the

Spectrum Analyzer RF Input.

NOTE

Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.

FM
Calibration
Procedure

The following steps in the procedure lets you cali-
brate the (1) FM Meter circuit; (2) FM Variable Gain
Linearity; (3) Locked, Locked Low-Noise, Unlocked
Narrow, and Unlocked Wide FM Mode Sensitivity;
and (4) FM Rear Panel Input Gain, and store the re-
sults in non-volatile memary (EEPROM) on the A17
CPU PCB.

NOTE

To ensure accurate calibration, each step of
this procedure must be performed in se-
quence.

1. Perform FM Meter calibration as follows:

a. At the

$

prompt on the PC screen, type:

calterm 123

and press .

b. Set up the Function Generator for a 100.0 kHz

sine wave with an output level of 0.707 volts
±0.01 volts RMS (2 volts peak to peak). Use a
frequency counter to verify the output fre-
quency of your function generator is set to
±1%. When done, press any key on the key-
board to continue calibration.

The

$

prompt will appear on the screen when

the calibration is complete.

c. Record step completion on the Test Record.

2. Perform FM Variable Gain Linearity calibration

as follows:

a. At the

$

prompt on the PC screen, type:

calterm 148

and press .

b. Set up the Function Generator for a +1.00 Vdc

output. When done, press any key on the key-
board to continue calibration.

The

$

prompt will appear on the screen when

the calibration is complete.

c. Record step completion on the Test Record.

682XXB/683XXB MM

4-45

FM

CALIBRATION

CALIBRATION

NOTE

To save the calibration data after
completing any calibration step,
type:

calterm 787

and press

.