General recommended information test equipment – Anritsu 682XXB User Manual
Page 23
682XXB/683XXB MM
1-13
INSTRUMENT
CRITICAL
SPECIFICATION
RECOMMENDED
MANUFACTURER/MODEL
USAGE
(1)
Modulation Analyzer
Frequency Input: 10 MHz
(or the IF of the Spectrum Analyzer)
AM Depth: 0% to 90%
AM Modulation Rates: DC to 100 kHz
Filters: 20 kHz lowpass, 300 Hz highpass
Hewlett-Packard, Model 8901A
P
Oscilloscope
Bandwidth: DC to 150 MHz
Vertical Sensitivity: 2mV/division
Horizontal Sensitivity: 50 ns/division
Tektronix, Inc. Model TAS485
P, T
Mixer
Frequency Range: 1 to 26 GHz
Macom Micro Electronics Div.
Model DMS1-26A
P
Scalar Network
Analyzer, with
RF Detector
Frequency Range: 0.01 to 50 GHz
ANRITSU, Model 562, with
RF Detector:
560-7K50 (0.01 to 40 GHz)
560-7VA50 (0.01 to 50 GHz)
SC5198 (40 to 60 GHz)
C, T
Adapter
K (male) to 2.4 mm (female)
Adapts the Power Sensor, HP 8487A, to
the 682XXB/683XXB RF OUTPUT con-
nector (
£
40 GHz models)
Hewlett-Packard
Part Number: HP 11904D
C, P
Adapter
Adapts the MP716A4 Power Sensor to
the ML4803A Power Meter
ANRITSU MA4002B
C, P
Adapter
WR15 to V (male)
Adapts the MP716A4 Power Sensor to
the 682XXB/683XXB RF OUTPUT
connector (>40 GHz models)
ANRITSU, Model 35WR15V
C, P
Attenuator
Frequency Range: DC to 40 GHz
Max Input Power: >+17 dBm
Attenuation: 10 dB
ANRITSU, Model 41KC-10
C, P
Attenuator
Frequency Range: DC to 40 GHz
Max Input Power: >+17 dBm
Attenuation: 20 dB
ANRITSU, Model 41KC-20
P
Attenuator
Frequency Range: DC to 60 GHz
Max Input Power: >+17 dBm
Attenuation: 10 dB
ANRITSU, Model 41V-10
C, P
Attenuator
Frequency Range: DC to 60 GHz
Max Input Power: >+17 dBm
Attenuation: 20 dB
ANRITSU, Model 41V-20
P
RF Detector
Frequency Range: 0.01 to 40 GHz
Output Polarity: Negative
ANRITSU, Model 75KC50
(K input/BNC output connectors)
T
RF Detector
Frequency Range: 0.01 to 50 GHz
Output Polarity: Negative
ANRITSU, Model 75VA50
(V input/BNC output connectors)
T
Table 1-2.
Recommended Test Equipment (2 of 3)
GENERAL
RECOMMENDED
INFORMATION
TEST EQUIPMENT