This procedure provides the steps necessary to perform ALC Slope
calibration. The ALC Slope is calibrated to adjust for decreasing out-
put power-vs-output frequency in full band analog sweep.
Equipment
Setup
Connect the equipment, shown in Figure 4-7, as fol-
lows:
1. Interface the PC to the 683XXB by performing
the initial setup procedure, pages 4-7 to 4-12.
2. Using the Auxiliary I/O cable, connect the
683XXB rear panel
AUX I/O
connector to the 562
Network Analyzer AUX I/O connector.
3. Using the GPIB cable, connect the 562 Network
Analyzer DEDICATED GPIB connector to the
683XXB
IEEE-488 GPIB
connector.
4. Connect the RF Detector to the 562 Network
Analyzer Channel A Input connector.
5. Connect the 683XXB
RF OUTPUT
connector to
the RF Detector via a 10 dB Attenuator.
682XXB/683XXB MM
4-27
ALC SLOPE
CALIBRATION
CALIBRATION
Serial
I/O
683XXB Signal Generator
COM1
or
COM2
IBM-Compatible PC
562 Network Analyzer
10 dB
Attenuator
GPIB
AUX
I/O
RF Detector
Dedicated
GPIB
RF
Output
A
AUX
I/O
Figure 4-7.
Equipment Setup for ALC Slope Calibration