4-12
AM CALIBRATION
This procedure provides the steps necessary to perform AM calibra-
tion. This consists of calibrating the AM Calibration DAC, the AM Me-
ter circuit, and the AM Function Generator. The AM Calibration DAC
is calibrated for input sensitivities of 100%/V (linear mode) and
25 dB/V (logarithmic mode) for frequencies
£2 GHz and >2 GHz.
Equipment
Setup
Connect the equipment, shown in Figure 4-10, as
follows:
1. Interface the PC to the 682XXB/683XXB by per-
forming the initial setup procedure, pages 4-7 to
4-12.
2. Connect the Function Generator Output to the
BNC tee. Connect one leg of the tee to the
682XXB/683XXB front panel
AM IN
. Connect the
other leg of the tee to the DMM input.
3. Calibrate the Power Meter with the Power Sen-
sor.
682XXB/683XXB MM
4-33
AM
CALIBRATION
CALIBRATION
RF
OUT
Power Meter
Power
Sensor
Serial
I/O
682XXB / 683XXB
COM1
or
COM2
IBM-Compatible PC
Function
Generator
OUTPUT
AM
IN
BNC
Tee
INPUT
DMM
Figure 4-10.
Equipment Setup for AM Calibration