Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual
Page 218
5-52
682XXB/683XXB MM
TROUBLESHOOTING
TROUBLESHOOTING
TABLES
Error 147 Internal FM circuitry failed
Description: Indicates a failure of the internal frequency modulation
function. The 682XXB/683XXB may or may not provide frequency
modulation of the RF output signal using modulating signals from an
external source.
Step 1.
Set up the 682XXB/683XXB as follows:
a.
Press
SYSTEM
, then
Reset
.
b.
Press
MODULATION
, then
FM
to go to the Internal FM
Status display.
c.
Press
On/Off
to turn internal frequency modulation on.
Step 2.
Using an oscilloscope, verify the presence of a 10 volt peak to
peak sine wave signal with a period of 100
ms at the rear
panel
FM OUT
connector.
G
If present, replace the A11 PCB.
G
If not present, replace the A8 PCB.
Table 5-26.
Error Message 147
Error 148 Pulse 40 MHz reference circuitry failed.
Description: Indicates a failure of the pulse generator 40 MHz oscilla-
tor circuitry. The pulse generator may still function; however, the
40 MHz oscillator is not phase locked to the 10 MHz reference time-
base. The pulse modulation function may or may not operate. Error 106
(Power Supply not Phase-locked) may also be displayed.
Step 1.
Disconnect cable W108 at A6J1.
Step 2.
Using an oscilloscope, verify the presence of a 10 MHz at the
end of cable W108.
G
If present, replace the A6 PCB.
G
If not present, go to step 3.
Step 3.
Reconnect cable W108 to A6J1 and disconnect it at A5J4.
Step 4.
Using the oscilloscope, verify the presence of a 10 MHz TTL
signal at A5J4.
G
If present, replace the cable W108.
G
If not present, go to step 5.
Table 5-27.
Error Message 148 (1 of 2)