Rise/fall time and overshoot 3-40, Rise/fall time and overshoot -40 – Anritsu 682XXB User Manual

Page 100

background image

Rise/Fall
Time and
Overshoot

The following procedure lets you measure the rise
time, fall time, and overshoot parameters of the
pulse modulation waveform.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to 5.0 GHz.

d. Press

Edit L1

to open the current power level

parameter for editing.

e. Set L1 to the maximum leveled output power

level for the instrument being tested (refer to
Table 3-2, page 3-6).

f. Press

MODULATION

, then

Pulse

. The Inter-

nal Pulse Status menu is displayed.

g. At the Internal Pulse Status menu, press

Edit Width

and set pulse width W1 to 5.0

ms.

h. Press

Edit Period

and set the PRI to 0.01 ms.

i. Press

On/Off

to turn pulse modulation on.

2. On the Oscilloscope, turn off CH.4. Observe the

output of the Pulse Detector on the Oscilloscope.
Refer to Figure 3-11 to interpret the detector
waveform. Adjust the Oscilloscope controls to
measure the following waveform parameters.

a. Rise/Fall Time (10 to 90%): <10 ns

b. Pulse Width: 5

ms ±10 ns (5 ns, typical)

c. PRI: 0.010 ms

±10 ns (5 ns, typical)

d. Overshoot: <10%

3. Record the results on the Test Record.

3-40

682XXB/683XXB MM

PERFORMANCE

PULSE MODULATION TESTS:

VERIFICATION

RISE TIME, FALL TIME, OVERSHOOT, AND LEVEL