Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual
Page 205
682XXB/683XXB MM
5-39
YIG-tuned Oscillator
Error 124 Full Band Unlocked and Unleveled
Error 125 8.4-20 GHz Unlocked and Unleveled
Error 126 2-8.4 GHz Unlocked and Unleveled
Description: These error messages indicate either a failure of one or
both of the oscillators in the 2-20 GHz YIG-tuned oscillator assembly.
Step 1.
Connect a 562 Scalar Network Analyzer to the 682XXB/
683XXB as follows:
a.
Connect the 682XXB/683XXB
AUX I/O
to the 562
AUX
I/O
.
b.
Connect the 562
DEDICATED GPIB
to the 682XXB/
683XXB
IEEE-488 GPIB
.
c.
Connect the RF Detector to the 562 Channel A Input.
Step 2.
Set up the 682XXB/683XXB as follows:
a.
682XXB Setup:
CW/SWEEP SELECT: Step
F1: 2.000 GHz
F2: 20.000 GHz
Number of Steps: 400
683XXB Setup:
CW/SWEEP SELECT: Analog
F1: 2.000 GHz
F2: 20.000 GHz
Sweep Time: 0.100 Sec
Step 3.
Set up the 562 Scalar Network Analyzer as follows:
a.
Press SYSTEM MENU key.
b.
From System Menu display, select RESET.
c.
Press CHANNEL 2 DISPLAY: OFF.
d.
Press CHANNEL 1 DISPLAY: ON.
e.
Press CHANNEL 1 MENU key.
f.
From the Channel 1 Menu display, select POWER.
Step 4.
Using the scalar network analyzer, measure the RF output
directly at the YIG-tuned oscillator’s output connector. The
amplitude of the RF signal should be >4 dBm throughout
the full sweep.
Table 5-18.
Error Messages 124, 125, and 126 (1 of 2)
TROUBLESHOOTING