Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual
Page 186
5-20
682XXB/683XXB MM
TROUBLESHOOTING
TROUBLESHOOTING
TABLES
Step 2.
Perform the following procedure to isolate malfunctions
when the voltages from several regulated supplies are
incorrect.
a.
Place the
LINE
switch to
STANDBY
.
b.
Replace the A15 PCB assembly.
c.
Place the
LINE
switch to
OPERATE
and measure the
regulated voltages per Table 5-7.
G
If the voltages are correct, the problem is cleared.
G
If the voltages are incorrect, go to step d.
d.
Place the
LINE
switch to
STANDBY
.
e.
Replace the A18 PCB.
f.
Place the
LINE
switch to
OPERATE
and measure the
regulated voltages per Table 5-7.
G
If the voltages are correct, the problem is cleared.
G
If the voltages are still incorrect, contact your local
ANRITSU
service center for assistance.
Step 3.
Run self-test again.
G
If no error message is displayed, the problem is cleared.
G
If error 105 displays again, contact your local
ANRITSU
service center for assistance.
Table 5-6.