beautypg.com

Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 213

background image

682XXB/683XXB MM

5-47

TROUBLESHOOTING

TROUBLESHOOTING

TABLES

Error 130 2-3.3 GH Switched Filter
Error 131 3.3-5.5 GH Switched Filter
Error 132 5.5-8.4 GH Switched Filter
Error 133 8.4-13.25 GH Switched Filter
Error 134 13.25-20 GH Switched Filter

Description: Each of these error messages indicates a failure in a
switched filter path within the switched filter assembly. The 682XXB/
683XXB may or may not produce an RF output in the frequency range of
the failed switched filter path.

Step 1.

Set up the 682XXB/683XXB as follows:

a.

682XXB Setup:
CW/SWEEP SELECT: Step
F1: 2.000 GHz
F2: 20.000 GHz
Number of Steps: 400

683XXB Setup:
CW/SWEEP SELECT: Analog
F1: 2.000 GHz
F2: 20.000 GHz
Sweep Time: 0.100 Sec

Step 2.

Connect the X input of an oscilloscope to the 682XXB/
683XXB rear panel

HORIZ OUT

connector.

Step 3.

Using the oscilloscope, check for the switched filter PIN
switch drive voltages at the test points shown in Table 5-21.

G

If the PIN switch drive voltages are correct, replace the
switched filter assembly.

G

If the PIN switch drive voltages are incorrect, replace the
A9 PCB.

Table 5-20.

Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (7 of 8)

Test Point

Active Frequency

Range

Active

Voltage

Inactive

Voltage

A9TP18

2 to 3.3 GHz

–2V

+1V

A9TP10

3.3 to 5.5 GHz

–2V

+1V

A9TP12

5.5 to 8.4 GHz

–2V

+1V

A9TP16

8.4 to 13.25 GHz

–2V

+1V

A9TP21

13.25 to 20 GHz

–2V

+1V

A9TP17

2 to 8.4 GHz

–2V

+2V

Table 5-21.

Switched Filter PIN Switch Drive Voltages

This manual is related to the following products: