11 alc bandwidth calibration 4-31, Equipment setup 4-31, Bandwidth calibration 4-31 – Anritsu 682XXB User Manual
Page 141: Alc bandwidth calibration, 4-31, Alc bandwidth calibration -31, Equipment setup -31 bandwidth calibration -31, Alc bandwidth calibration calibration
4-11
ALC BANDWIDTH
CALIBRATION
This procedure provides the steps necessary to perform ALC Band-
width calibration. The ALC Bandwidth is adjusted to compensate for
gain variations of the modulator. The adjustment is performed for
each frequency band. This provides a more consistent bandwith
throughout the frequency range of the instrument.
Equipment
Setup
Connect the equipment, shown in Figure 4-9, as fol-
lows:
1. Interface the PC to the 682XXB/683XXB by per-
forming the initial setup procedure, pages 4-7 to
4-12.
NOTE
Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.
Bandwidth
Calibration
The following procedure lets you (1) calibrate the
ALC bandwith and (2) store the calibration data in
non-volatile memory (EEPROMs) on the A17 CPU
PCB.
1. Enter the ALC Bandwidth calibration routine as
follows:
a. At the
$
prompt on the PC display, type:
calterm 110
and press
The following will appear on the PC display:
Enter calculated gain X.XX
Input:
b. Type:
1.50
and press
682XXB/683XXB MM
4-31
ALC BANDWIDTH
CALIBRATION
CALIBRATION
Serial
I/O
682XXB / 683XXB
COM1
or
COM2
IBM-Compatible PC
Figure 4-9.
Equipment Setup for AM Bandwidth Calibration