TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Table 6-7: RS-232 Command Summary

Commands

Definition

?

Print help screen

T SET ALL

Enable display of all test variables during T LIST

T SET name

Display only NAME during T LIST

T LIST [ALL]

Print all test variables enabled with T SET or ALL
warnings

T name

Print single test, "name" from Table 6-9

T CLEAR ALL

Disable T LIST, use with T SET name

W SET ALL

Enable display of all warnings during W LIST

W LIST [ALL]

Print warnings enabled with W SET or ALL warnings

W name

Print individual "name" warning from Table 6-10

W CLEAR ALL

Disable W LIST, use with W SET

C command

Execute calibration "command" from Table 6-12

D LIST

Prints all I/O signal values

D name

Prints single I/O signal value/state

D name=value

Sets variable to new "value"

D LIST NAMES

Lists diagnostic test names

D ENTER name

Enters and starts 'name' diagnostic test

D EXIT

Exits diagnostic mode

D RESET

Resets analyzer(same as power-on)

D RESET RAM

System reset, plus erases RAM. Initializes DAS, SO

2

conc

readings, calibration data not affected

D RESET EEPROM

System reset, plus erases EEPROM (RAM_RESET
actions + setup variables, calibration to default values)

V LIST

Print all easy variable names from Table 9-5

V name

Print individual "name" variable

V name=value

Sets variable to new "value"

V CONFIG

Print analyzer configuration