Able, Rs-232 c, Ommand – Teledyne 6200A - Sulfides analyzer User Manual
Page 95: Ummary

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Table 6-7: RS-232 Command Summary
Commands
Definition
?
Print help screen
T SET ALL
Enable display of all test variables during T LIST
T SET name
Display only NAME during T LIST
T LIST [ALL]
Print all test variables enabled with T SET or ALL
warnings
T name
Print single test, "name" from Table 6-9
T CLEAR ALL
Disable T LIST, use with T SET name
W SET ALL
Enable display of all warnings during W LIST
W LIST [ALL]
Print warnings enabled with W SET or ALL warnings
W name
Print individual "name" warning from Table 6-10
W CLEAR ALL
Disable W LIST, use with W SET
C command
Execute calibration "command" from Table 6-12
D LIST
Prints all I/O signal values
D name
Prints single I/O signal value/state
D name=value
Sets variable to new "value"
D LIST NAMES
Lists diagnostic test names
D ENTER name
Enters and starts 'name' diagnostic test
D EXIT
Exits diagnostic mode
D RESET
Resets analyzer(same as power-on)
D RESET RAM
System reset, plus erases RAM. Initializes DAS, SO
2
conc
readings, calibration data not affected
D RESET EEPROM
System reset, plus erases EEPROM (RAM_RESET
actions + setup variables, calibration to default values)
V LIST
Print all easy variable names from Table 9-5
V name
Print individual "name" variable
V name=value
Sets variable to new "value"
V CONFIG
Print analyzer configuration
6-14