Specifications, agency approvals, warranty, Specifications, 3 specifications, agency approvals, warranty -1 – Teledyne 6200A - Sulfides analyzer User Manual
Page 52: Pecifications, 3 specifications, agency approvals, warranty, 1 specifications

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
3 SPECIFICATIONS, AGENCY APPROVALS,
WARRANTY
3.1 Specifications
Ranges
In 1ppb increments from 50ppb to 20,000ppb, dual ranges or
autoranging
Noise at Zero
1
0.2 ppb RMS
Noise at Span
1
<0.5% of reading (above 50 ppb)
Lower Detectable Limit
2
0.4 ppb RMS
Zero Drift
3
<0.5 ppb/24 hours
Zero Drift
3
1 ppb/7 days
Span Drift
3
<0.5%
FS/24hrs
Lag Time
20 sec
Rise Time
95% in <120 sec
Fall Time
95% in <100 sec
Sample Flow Rate
700cc/min. ±10%
Linearity
1% of full scale
Precision
0.5% of reading
1
Temperature Range
5-40
o
C
Temp Coefficient
< 0.1% per
o
C
Humidity
0 - 95% RH, non-condensing
Voltage Coefficient
< 0.05% per V
Dimensions HxWxD
7"x17" x 23.6" (18cm x 43cm x 61cm)
Weight, Analyzer
45 lbs (20.5 kg) w/internal pump
Power, Analyzer
110v/60Hz, 220v/50Hz, 240v/50Hz, 250 watts
Power, Analyzer
4
230v/50Hz,
2.5A
Environmental Conditions
Installation Category (Overvoltage Category) II
Pollution Degree 2
Recorder Output
5
0-100 mV, 0-1, 5, 10v ; resolution of 1 part in 1024 of selected
voltage or current range
Status Option
12 Status Outputs from opto-isolator
Measurement Units
ppb, ppm, ug/m
3
, mg/m
3
1. As defined by USEPA.
2. Defined as twice the zero noise level.
3. At constant temperature and voltage.
4. Electrical ratings for CE Mark compliance
5. Bi-polar.
3-1