beautypg.com

Specifications, agency approvals, warranty, Specifications, 3 specifications, agency approvals, warranty -1 – Teledyne 6200A - Sulfides analyzer User Manual

Page 52: Pecifications, 3 specifications, agency approvals, warranty, 1 specifications

background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

3 SPECIFICATIONS, AGENCY APPROVALS,

WARRANTY

3.1 Specifications

Ranges

In 1ppb increments from 50ppb to 20,000ppb, dual ranges or
autoranging

Noise at Zero

1

0.2 ppb RMS

Noise at Span

1

<0.5% of reading (above 50 ppb)

Lower Detectable Limit

2

0.4 ppb RMS

Zero Drift

3

<0.5 ppb/24 hours

Zero Drift

3

1 ppb/7 days

Span Drift

3

<0.5%

FS/24hrs

Lag Time

20 sec

Rise Time

95% in <120 sec

Fall Time

95% in <100 sec

Sample Flow Rate

700cc/min. ±10%

Linearity

1% of full scale

Precision

0.5% of reading

1

Temperature Range

5-40

o

C

Temp Coefficient

< 0.1% per

o

C

Humidity

0 - 95% RH, non-condensing

Voltage Coefficient

< 0.05% per V

Dimensions HxWxD

7"x17" x 23.6" (18cm x 43cm x 61cm)

Weight, Analyzer

45 lbs (20.5 kg) w/internal pump

Power, Analyzer

110v/60Hz, 220v/50Hz, 240v/50Hz, 250 watts

Power, Analyzer

4

230v/50Hz,

2.5A

Environmental Conditions

Installation Category (Overvoltage Category) II
Pollution Degree 2

Recorder Output

5

0-100 mV, 0-1, 5, 10v ; resolution of 1 part in 1024 of selected
voltage or current range

Status Option

12 Status Outputs from opto-isolator

Measurement Units

ppb, ppm, ug/m

3

, mg/m

3

1. As defined by USEPA.
2. Defined as twice the zero noise level.
3. At constant temperature and voltage.
4. Electrical ratings for CE Mark compliance
5. Bi-polar.

3-1