Variables commands and messages, 8 variables commands and messages -24, Able – Teledyne 6200A - Sulfides analyzer User Manual

Page 105: 16: rs-232 o, Perating, Odes, 8 variables commands and messages

background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

6.8.8 VARIABLES Commands and Messages

Table 6-16: RS-232 Operating Modes

Mode

Description

ZERO CAL A

Automatic zero calibration

ZERO CAL R

Remote zero calibration

ZERO CAL M

Manual zero calibration

SPAN CAL A

Automatic span calibration

SPAN CAL R

Remote span calibration

SPAN CAL M

Manual span calibration

M-P CAL

Manual multi-point calibration

DIAG ELEC

Electrical diagnostic test

DIAG OPTIC

Optical diagnostic test

DIAG AOUT

D/A output diagnostic test

DIAG

Main diagnostic menu

DIAG I/O

Signal I/O diagnostic

DIAG RS232

RS232 output diagnostic

DIAG ERASE

Memory erase diagnostic

SETUP x.x

Setup mode (x.x is software version)

SAMPLE A

Sampling; automatic cal. enabled

SAMPLE

Sampling; automatic cal. disabled

The 6200A operational modes are listed above. To list the analyzer's current mode type:

V MODE

Lists 6200A current operational mode

See Table 6-14 for list of operational modes and meanings.

Model 6200A Internal Variables

The 6200A has a number of internal setup variables. Essentially all of these are set at time of
manufacture and should not need to be changed in the field. A list of user accessible variables is
shown in Table 9-5.