beautypg.com

Teledyne 6200A - Sulfides analyzer User Manual

Page 174

background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Like the display, the overall functioning of the CPU can be confirmed by a simple test.

1. Locate the CPU board on the mother board by referring to Figure 2-5.

2. Power the instrument on.

3. Locate the red LED at the top left edge of the board.

4. It should be flashing at a frequency of about once per second.

5. This flashing indicates the board is powered up and is executing instructions.

RS-232 diagnostic procedures are described in Section 9.3.2.2. It is possible for the UART driver
chip to malfunction in either or both of the input or output ports.

9-31