Able, 2: s, Etup – Teledyne 6200A - Sulfides analyzer User Manual
Page 85: Utomatic, Alibration, Setup described in table 6-2 operation of

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Table 6-2: Setup Automatic Zero/Span Calibration
Default
Description
Timer Enable
ON
Enable or disable automatic calibration timer.
Starting Date
01-JAN-95
MM:DD:YY
Starting Time
00:00
HH:MM, 0 - 23 hours and 0 - 59 minutes
Delta Days
1
Delay days between each calibration (0 - 365 days)
Delta Time
00:00
HH:MM, 0 - 23 hours and 0 - 59 minutes
Duration
15.0 minutes
1 - 60.0 minutes
Calibrate
OFF
on/off. If ON is selected, it will adjust the calibration.
The Timer Enable can be set to “OFF” to disable the automatic calibration timer while the
remote RS-232 calibration of specific sequence can be initiated.
NOTE
Avoid setting two or more sequences at the same time of the day. Any new
sequence which is initiated whether from a timer, the RS-232, or the contact
closure inputs will override any sequence which is in progress.
The programmed start time must be a minimum of
5 minutes later than the real time clock.
Examples of possible sequences are as following under any one of three available SEQx.
Example 1: to perform zero-span calibration check once per day at 10:30 pm, 12/20/98.
1) MODE: ZERO-HI
2) TIMER ENABLE: ON
3) STARTING DATE: 12/20/98
4) STARTING TIME: 22:30
5) DELTA DAYS: 1
6) DELTA TIME: 00:00
7) DURATION: 15.0 MINUTES
8) CALIBRATE: OFF
6-4