Analyzer Operator Manual, 02164, Rev. G

Table 6-2: Setup Automatic Zero/Span Calibration

Default

Description

Timer Enable

ON

Enable or disable automatic calibration timer.

Starting Date

01-JAN-95

MM:DD:YY

Starting Time

00:00

HH:MM, 0 - 23 hours and 0 - 59 minutes

Delta Days

1

Delay days between each calibration (0 - 365 days)

Delta Time

00:00

HH:MM, 0 - 23 hours and 0 - 59 minutes

Duration

15.0 minutes

1 - 60.0 minutes

Calibrate

OFF

on/off. If ON is selected, it will adjust the calibration.

The Timer Enable can be set to “OFF” to disable the automatic calibration timer while the
remote RS-232 calibration of specific sequence can be initiated.

NOTE

Avoid setting two or more sequences at the same time of the day. Any new

sequence which is initiated whether from a timer, the RS-232, or the contact

closure inputs will override any sequence which is in progress.

The programmed start time must be a minimum of

5 minutes later than the real time clock.

Examples of possible sequences are as following under any one of three available SEQx.

Example 1: to perform zero-span calibration check once per day at 10:30 pm, 12/20/98.

1) MODE: ZERO-HI

2) TIMER ENABLE: ON

3) STARTING DATE: 12/20/98

4) STARTING TIME: 22:30

5) DELTA DAYS: 1

6) DELTA TIME: 00:00

7) DURATION: 15.0 MINUTES

8) CALIBRATE: OFF

6-4