beautypg.com

Able, 1: f, Inal – Teledyne 6200A - Sulfides analyzer User Manual

Page 50: Est and, Alibration, Alues, Ontinued

background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Table 2-1: Final Test and Calibration Values (Continued)

Span and Cal Values

Parameter

Observed
Value

Units

Nominal Range

Reference Section

SO

2

Span Conc

PPB

20 - 20,000

Table 7-3

SO

2

Slope

1.0 ± 0.3

7.10

SO

2

Offset

mV

<250

7.10

Noise at Zero (rms)

PPB

0.1 - 0.2

Table 9-1

Noise at Span (rms)

PPB

0.5% of reading
(above 50 ppb)

Table 9-1

Measured Flows

Parameter

Observed
Value

Units

Nominal Range

Reference Section

Sample Flow

cc/min

700 ± 10%

9.2.2, 9.3.6,
Figure 9-6

IZS Purge Flow

cc/min

50 ± 10

6.3

Factory Installed Options

Option Installed

Power Voltage/Frequency

Rack Mount, w/ Slides

Rack Mount, w/ Ears Only

Fluorocarbon Zero/Span Valves

Internal Zero/Span - IZS

Permeation Tube (Output Specification)

External Pump

4-20 mA Current Loop Output

REC

DAS

SO

2

(RANGE 1)

SO

2

(RANGE 2)

SPARE

TEST OUTPUT

PROM #

Serial #

Date

Technician

2-13