beautypg.com

Command summary, 2 command summary -13, Able – Teledyne 6200A - Sulfides analyzer User Manual

Page 94: 6: rs-232 t, Erminal, Diting, 2 command summary

background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Table 6-6: RS-232 Terminal Mode Editing Keys

Key

Function

CR (carriage return)

Execute command

BS (backspace)

Backspace one character to the left

ESC (escape)

Erase entire line

Control-R (ASCII 18 decimal)

Recall previous command

Control-E (ASCII 5 decimal)

Recall and execute previous command


Commands are not case-sensitive; you should separate all command elements (i.e. keywords, data
values, etc.) by spaces.

Words such as T, SET, LIST, etc. are called keywords and are shown on the help screen in
uppercase, but they are not case-sensitive. You must type the entire keyword; abbreviations are not
accepted.

OBTAINING HELP

Typing "?" followed by Return or Enter will cause a help screen to be displayed.

6.8.2 Command Summary

The information contained in the rest of this section covers all of the normal commands that are
required to operate the instrument from a remote terminal. If you are going to be writing computer
programs to communicate with the 6200A (i.e., operating the port in COMPUTER MODE) we
suggest that you order a supplementary manual "The RS-232 Interface". This manual shows
additional features of the port designed to support a computer driven interface program.

6-13