TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Table 9-5: Model 6200A Variables

No.

Name

Units

Default
Value

Value
Range

Description

0

DAS_HOLD_OFF

MIN

15.0

0.5-20.0

Hold off duration after
calibration or diagnostic
mode.

1 TPC_ENABLE

ON

ON-OFF Temp/Pres

compensation

enable

2 RCELL_SET

o

C

50

30-70

Reaction cell temperature set
point

3 IZS_SET

°C

50

30-70

IZS temperature set point

4

DYN_ZERO

OFF

OFF/ON

Enable to adjust zero
calibration through remote
contact closure

5

DYN_SPAN

OFF

OFF/ON

Enable to adjust span
calibration through remote
contact closure

6

RS232_MODE

Bit Field

8

0-99999

Value is SUM of following
decimal numbers:

1=

enable quiet mode

2=

enable computer mode

3=

enable security feature

4=

enable front panel RS-

232 menus

7

CLOCK_ADJ

Sec.

0

± 60

Real-time clock adjustment

9-19