Das commands and reports, 7 das commands and reports -21, Able – Teledyne 6200A - Sulfides analyzer User Manual
Page 102: Rs-232 d, Iagnostic, Ommand, Ummary, 15: rs-232 das c, Ommands, Able 6-14

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
The following is a summary of the Diagnostic commands.
Table 6-14: RS-232 Diagnostic Command Summary
Command
Description
D LIST
Prints all I/O signal values. See Table 9-4 for Sig I/O definitions.
D name=value
Examines or sets I/O signal. For a list of signal names see Table 9-4
in Section 9. Must issue D ENTER SIG command before using this
command.
D LIST NAMES
Prints names of all diagnostic tests
D ENTER SIG
D ENTER OT
D ENTER ET
D ENTER TASK
Executes SIGNAL I/O diagnostic test.
Executes Optic Test diagnostic test.
Executes Elect Test diagnostic test.
Displays a listing of the tasks and their status.
Use D EXIT to leave these diagnostic modes.
D EXIT
Must use this command to exit SIG, ET or OT Diagnostic modes
D RESET
Resets analyzer software. (Same as power on.)
D RESET RAM
Resets analyzer software and erases RAM. Erases SO
2
conc values.
Keeps setup variables and calibration. (Same as installing new
software version.)
D RESET EEPROM
Resets analyzer software and erases RAM and EEPROM. Returns all
setup variables to factory defaults, resets calibration value.
6.8.7 DAS Commands and Reports
RS-232 Commands
In addition to accessing the data acquisition system and the stored data from the instrument front
panel, you can also access the data acquisition and the stored data from the RS-232 interface.
There are two RS-232 commands, listed in the table below.
Table 6-15: RS-232 DAS Commands
RS-232 DAS Commands
Command
Description
D [id] PRINT [“name”]
Prints specified data channel’s properties
D [id] REPORT “name” [RECORDS=number]
[COMPACT|VERBOSE]
Prints the data for the specified data channel
6-21