Table 2-1: Final Test and Calibration Values

TEST Values

Observed
Value

Units

Nominal Range

Reference Section

RANGE

PPB

50 - 20,000

5.3.4

STABIL

PPB

0.1 - 2

5.2.1, 9.1.1,
Table 9-1, 9.2.5

PRESS

in-Hg

25 - 35

9.1.1, 9.3.6,
Table 9-1

SAMP FL

CC / MIN

650 ± 10%

9.2.2, 9.3.6,
Table 9-1

PMT

mV

0 - 5000

9.4.1

UV LAMP

mV

2000 - 4000 typical

9.4.2

STR. LGT

PPB

<100

Table 9-1

DRK PMT

mV

-50 - +200

9.4.3, Table 9-1

DRK LMP

mV

-50 - +200

Table 9-1

SLOPE

1.0 ± 0.3

7.10

OFFSET mV

<250

7.10

HVPS

V

550 - 900 constant

9.3.10

DCPS

mV

2500 ± 200

9.3.5

RCELL TEMP

o

C

50 ± 1

9.3.7

BOX TEMP

o

C 8-50

9.3.4.1

PMT TEMP

o

C

7 ± 1

9.3.9

IZS TEMP

o

C

50 ± 0.3

9.5.4

Electric Test & Optic Test

Electric Test

PMT Volts

mV

2000 ± 1000

9.1.3.2

SO

2

Conc

PPB

1000 ± 500

9.1.3.2

Optic

Test

PMT Volts

mV

2000 ± 1000

9.1.3.3

SO

2

Conc

PPB

1000 ± 500

9.1.3.3

(table continued)

2-12