Table 7-1: Types of Zero/Span Checks and Calibrations

Section

Type of Cal or Check

Description

7.1

Manual Z/S Check - Cal gas
through sample port

This calibration option expects the calibration gas
coming in through the sample port. IZS and
Zero/Span valves do not operate.

7.2

Manual Z/S Check or Calibration
with Z/S valve Option

How to operate Zero/Span Valve Option.
Can be used to check or adjust calibration.

7.3

Manual Z/S Check or Calibration
with IZS Option

How to operate IZS Option. Can be used to check or
adjust calibration.

7.4

Automatic Z/S Check with Z/S
Valves or IZS Option

Operates Z/S valves or IZS once per day to check
the calibration.

7.5

Dynamic Z/S Calibration with Z/S
Valves or IZS Option

Operates Z/S valves or IZS once per day and adjusts
the calibration.

7.6

Calibrate using SO

2

Permeation

Tube

Allows calibration using SO

2

gas from the IZS

permeation tube.

7.7

Use of Z/S Valve or IZS with
Remote Contact Closure

Operates Z/S valves or IZS with rear panel contact
closures. Without valves or IZS, can be used to
switch instrument into zero or span cal mode. Used
for either checking or adjusting zero/span.

7.8

EPA Protocol Calibration

Covers methods to be used if data is for EPA
SLAMS monitoring.

7.9

Special calibration requirements
for Dual Range or Auto Range

Covers special requirements if using Dual Range or
Auto Range.

7.10

Calibration Quality

Information on how to determine if the calibration
performed will result in optimum instrument
performance.

7.11

References

Contains a list of references on quality control and
calibration.

7-2