Teledyne 6200A - Sulfides analyzer User Manual

Page 44

background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

The next task is to calibrate the analyzer. There are several ways to do a calibration; they are
summarized in Table 7-1. For a preliminary checkout we recommend calibration with zero air and
span gas coming in through the sample port. The procedure is:

Step 1 - Enter the expected SO

2

span gas concentration

Step Number Action

Comment

1.

Press CAL-CONC

This key sequence causes the 6200A to prompt for the
expected SO

2

concentration. Enter the SO

2

span concentration

value by pressing the key under each digit until the expected
value is set.

2.

Press ENTR

ENTR stores the expected SO

2

span value. This value will be

used in the internal formulas to compute subsequent SO

2

concentration values.

3.

Press EXIT

Returns instrument to SAMPLE mode.

4. Press

SETUP-

RNGE-SET

If necessary you may want to change ranges. Normally the
instrument is shipped with range set at 500 ppb.

5.

Press EXIT

Returns the instrument to SAMPLE mode.

2-7