Teledyne 6200A - Sulfides analyzer User Manual
Page 20

TAI - Thermal Converter Operator Manual - Page 16
Table 2.4 – Temperature Controller - Initial Settings
PRESS
DISPLAY
INITIAL VALUE
SEL
P PROP BAND
UP/DOWN SET TO “11”
SEL
i INTEGRAL
SET TO “10”
SEL
d DERIVATIVE TIME
SET TO “7.7”
SEL
AL LOW ALARM
SETPOINT
SET TO “50” (C BELOW FINAL SETPOINT)
SEL
AH HIHG ALARM
SETPOINT
SET TO “50” (C ABOVE FINAL SETPOINT)
SEL
7C CYCLE TIME
SET TO “2”
SEL
HYS HYSTERESIS
SET TO “3”
SEL
A7 AUTOTUNE
SET TO “0” (OFF)
SEL
LOC LOCK
“0” (OPEN) “1” (LOCKED)
“2” (SV ONLY OPEN)
Table 2.4 shows a Secondary Menu of parameters that set more basic parameters of the controller, these
include the thermocouple type, the temperature units etc.
Table 2.5 – Temperature Controller - Secondary Menu
PRESS
DISPLAY
PARAMETER VALUE
SEL
HOLD TILL p-n1
SET TO “0”
SEL
p-n2
SET TO “3” (TYPE K THERMOCOUPLE)
SEL
p-dF DIGITAL FILTER
SET TO “5”
SEL
P-SL LOWER LIMIT
SET TO “32” (32C)
SEL
P-SU UPPER LIMIT
SET TO “1050” (1050C)
SEL
P-AL ALARM TYPE2
SET TO “900”
SEL
P-AH ALARM TYPE 1
SET TO “1050”
SEL
P-An HYTERESIS
SET TO “3”
SEL
P-dP DECIMAL LOCATION
SET TO “0”
SEL
PVOF PROCESS OFFSET
LEAVE AT “0”
SEL
SVOF SET POINT OFFSET
LEAVE AT “0”
SEL
P-F
SET TO “ºC” (CENTIGRADE)
SEL
FUZZY FUZZY LOGIC
SET TO “ON”
- 1220 - Multipoint flammable gas and vapor detection system (50 pages)
- 212R - Thermal conductivity analyzer (28 pages)
- 235 - Thermal conductivity analyzer (38 pages)
- 275R - Portable turbine generator purge gas analyzer (21 pages)
- 2000A-EU - General purpose thermal conductivity analyzer (86 pages)
- 2000XTC - Thermal conductivity analyzer (40 pages)
- 2010A - Split architecture thermal conductivity analyzer (110 pages)
- 2010B - Split architecture thermal conductivity analyzer (98 pages)
- 2020 - Explosion proof thermal conductivity analyzer (80 pages)
- 2120 - Trace Nitrogen in Argon Analyzer (66 pages)
- 2120XL - Trace Nitrogen Analyzer (85 pages)
- 2230R - Process Hydrogen Analyzer (26 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 3rd generation (updated 1/31/11) (30 pages)
- 2240 - Portable Handheld Hydrogen Leak Detector, 3rd generation (revision 2/29/08) (40 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 2nd generation (13 pages)
- 2750 - Portable turbine generator gas analzyer (40 pages)
- 300P - Percent oxygen analyzer (24 pages)
- 306WA - Analog trace oxygen analyzer (46 pages)
- 311 - Portable trace oxygen analyzer (19 pages)
- 311D - Portable trace oxygen analyzer with digital meter (18 pages)
- 311XL - Portable trace oxygen analyzer (18 pages)
- 316RA / RB / RAD / RBD - Oxygen analyzers (24 pages)
- 319R - Oxygen analyzer (23 pages)
- 320 Series - Portable oxygen detectors (24 pages)
- 326, 327 and 328 - Oxygen analyzers (45 pages)
- 329R - Oxygen analyzer (22 pages)
- 335 - Analog control room monitor for personnel safety (24 pages)
- 356WA - Analog trace oxygen analyzer (42 pages)
- 3000MA - Paramagnetic oxygen analyzer (63 pages)
- 3000MA - Paramagnetic oxygen analyzer Addendum (2 pages)
- 3000MB - Paramagnetic oxygen analyzer (59 pages)
- 3000PA - General purpose percent oxygen analyzer (69 pages)
- 3000PAEU - General purpose percent oxygen analyzer (78 pages)
- 3000PB - Bulkhead mount percent oxygen analyzer (82 pages)
- 3000TA - General purpose trace oxygen analyzer (75 pages)
- 3000TA-EU - General purpose trace oxygen analyzer (89 pages)
- 3000TA-XLEU - Trace oxygen analyzer (108 pages)
- 3000TB - Bulkhead mount trace oxygen analyzer (78 pages)
- 3000TB-XL - Trace oxygen analyzer (78 pages)
- 3000ZA - Trace oxygen analyzer (81 pages)
- 3000ZA-3X - Trace oxygen analyzer (72 pages)
- 3000ZA2G - Zirconium oxide analyzer (72 pages)
- 3000 Ultra Trace - PPB oxygen analyzer (72 pages)
- 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range (88 pages)
- 3010MA – Paramagnetic oxygen analyzer, no 0-100% range – (superceded) (88 pages)