TAI - Thermal Converter Operator Manual - Page 16

Table 2.4 – Temperature Controller - Initial Settings

PRESS

DISPLAY

INITIAL VALUE

SEL

P PROP BAND

UP/DOWN SET TO “11”

SEL

i INTEGRAL

SET TO “10”

SEL

d DERIVATIVE TIME

SET TO “7.7”

SEL

AL LOW ALARM
SETPOINT

SET TO “50” (C BELOW FINAL SETPOINT)

SEL

AH HIHG ALARM
SETPOINT

SET TO “50” (C ABOVE FINAL SETPOINT)

SEL

7C CYCLE TIME

SET TO “2”

SEL

HYS HYSTERESIS

SET TO “3”

SEL

A7 AUTOTUNE

SET TO “0” (OFF)

SEL

LOC LOCK

“0” (OPEN) “1” (LOCKED)

“2” (SV ONLY OPEN)

Table 2.4 shows a Secondary Menu of parameters that set more basic parameters of the controller, these
include the thermocouple type, the temperature units etc.

Table 2.5 – Temperature Controller - Secondary Menu

PRESS

DISPLAY

PARAMETER VALUE

SEL

HOLD TILL p-n1

SET TO “0”

SEL

p-n2

SET TO “3” (TYPE K THERMOCOUPLE)

SEL

p-dF DIGITAL FILTER

SET TO “5”

SEL

P-SL LOWER LIMIT

SET TO “32” (32C)

SEL

P-SU UPPER LIMIT

SET TO “1050” (1050C)

SEL

P-AL ALARM TYPE2

SET TO “900”

SEL

P-AH ALARM TYPE 1

SET TO “1050”

SEL

P-An HYTERESIS

SET TO “3”

SEL

P-dP DECIMAL LOCATION

SET TO “0”

SEL

PVOF PROCESS OFFSET

LEAVE AT “0”

SEL

SVOF SET POINT OFFSET

LEAVE AT “0”

SEL

P-F

SET TO “ºC” (CENTIGRADE)

SEL

FUZZY FUZZY LOGIC

SET TO “ON”