Luminex MAGPIX Hardware and Installation Manual User Manual
Page 74

1.
Execute STOP if a plate is running. Refer to the software manual for
instructions.
2. Turn off MAGPIX and unplug the power cord.
3. Remove the sample probe.
a. Open the side access door of MAGPIX.
b. Unscrew the probe fitting on top of the probe completely.
c. Grasp the probe gently and push up.
d. Lift the probe out of the top of the probe holder.
FIGURE 52.
Probe Assembly
1. Probe fitting - Unscrew and remove in step 3.
2. Probe holder
3. Probe - Push up gently and lift out of holder in step 3.
4. Clean the sample probe using a bath sonicator, a 10 mL syringe, or both. If you are using
a bath sonicator, place the tip of the sample probe in the bath sonicator for 2 to 5
minutes. If you are using a syringe, force distilled water through the tip of the sample
probe to its large end. This dislodges any debris clogging the tip.
Operational and Maintenance Procedures
61