8150 Flow Totalizer

9. AutoCAL

Calibration

The AutoCAL feature allows theTotalizer to be adjusted to match the fl ow rate to any external reference.
• Flow in the pipe should be as stable as possible for best results.
• If the fl ow rate display is erratic, set the SPEED (section 14) to 120 seconds during the AutoCAL procedure.
• The timebase on the reference meter must be the same as the 8150 Totalizer.

Example: The

Totalizer

fl ow rate shows 60 GPM, while an external reference indicates a true fl ow rate of 70 GPM.

Change

the

fl ow rate from 60 GPM to 70 GPM using AutoCAL.

1. Hold the ENTER key for 2 seconds, then enter the SECURITY CODE.
AutoCAL

is

the

fi rst item in the menu.

2. Press the ► key to select the AutoCAL function.
The display will show AutoCAL and the current fl ow rate will be fl ashing.
3. Press the ► key again to change the fl ow rate. The display shows "Set Flo" and the fi rst digit of the fl ow rate will begin fl ashing.
4. Press the ► key to advance the fl ashing element to the "6".
5. Press the ▲ key one time to change the "6" to "7".
6. Press the ENTER key to complete the automatic calibration process.
The display shows a new K-Factor with the fi rst digit fl ashing. This K-Factor is based on the change in fl ow rate.
7. Press the ENTER key again to accept the new value.
NOTE: If the display shows "ERR SetFlo" the procedure was unsuccessful because the calculated K-Factor is less than 0.001 or

greater than 999999. Verify the fl ow rate and start the AutoCAL procedure from step 1.

8. Press ▲ and ▼ keys together to store the new value in the memory.
The display will show "storing" for a few seconds, then return to normal operation.

8

7

2

3

5

6

4

1

2 s

Enter security code