GF Signet 8150 Battery-Powered Flow Totalizer User Manual
Page 5
5
8150 Flow Totalizer
5 sec.
6.2 Standard Menu Settings
Totalizers are shipped from the factory with these standard settings:
Function: Factory
set:
Description:
AUTO CALIBRATION
No setting; See section 9 for detailed information.
FLOW K-FACTOR
60
Number of sensor pulses per volumetric unit; Refer to sensor manual.
TOTAL K-FACTOR
1
Set the number of volumetric units per totalizer count; see section 11.
TIMEBASE
Minutes
Select
fl ow rate in seconds, minutes, hours or days. Section 12.
DECIMAL
XXX.X
Set the maximum decimal resolution. Section 13.
SPEED
30 s
Zero to 120 seconds averaging stabilizes readings in erratic fl ow conditions. Section 14.
SENSITIVITY
6
Momentarily overrides SPEED when fl ow rate changes signifi cantly. Section 15.
SECURITY CODE
0-0-0-0
Set a private code to prevent tampering. Section 7.
DEFAULT TOTALIZER
Permanent
Select from three totalizer options. Section 8.
TOTALIZER #2 RESET
Reset Totalizer #2 after entering the security code. Section 8.3.
6. Operation
6.1 No Flow and ELAPSED TIME Display
If the fl ow stops, the Flow Rate displays the number of hours
since fl ow was last detected. This display will alternate with the
normal Totalized Value and standard "no FLo" display every
fi ve seconds.
Any movement of the rotor in the pipe will reset the ELAPSED
TIME display.
The 8150-1 display shows the fl ow rate in large numerals and
a totalizer value in smaller numerals. Any one of three different
totalizers can be selected as the standard display (See section 8
for detailed information on the totalizers).
Press the ▲ or ▼ keys to scroll through all three totalizer values
during normal operation.
Illustrated: No fl ow for 3 hours