Burkert Type 8056 User Manual
Page 57

- 57 -
RS232_485_ETP_MODBUS_BU_REV02.doc
Read command: CFFKZ?
Value of the coefficient if accepted
5:ACCESS ERR if insufficient access
level
CFFKZ (CoeFFicient KZ) Reads or
sets the value of the zero coefficient
KZ
Help command: CFFKZ=?
min <> max
if accepted
5:ACCESS ERR if insufficient access
level
Read command: ICALE?
Status of cal. error recognition if
accepted
5:ACCESS ERR if insufficient access
level
Set command: ICALE=
n
0:OK if parameters accepted
2:PARAM ERR if
n
out of range
5:ACCESS ERR if insufficient access
level
ICALE (Ignore CALibration Errors)
Enables or disables the calibration
errors recognition.
Help command: ICALE=?
List of
num:description
if accepted
5:ACCESS ERR if insufficient access
level
Read command: C1CP1?
Value of the cal. point if accepted
1:CMD ERR if current output not
enabled
5:ACCESS ERR if insufficient access
level
C1CP1
(Current output 1
Calibration Point 1) Reads or sets
the value of the current output
calibration point 1 (4 mA).
Help command: C1CP1=?
min <> max
if accepted
1:CMD ERR if current output not
enabled
5:ACCESS ERR if insufficient access
level
Read command: C1CP2?
Value of the cal. point if accepted
1:CMD ERR if current output not
enabled
5:ACCESS ERR if insufficient access
level
C1CP2
(Current output 1
Calibration Point 2) Reads or sets
the value of the current output
calibration point 2 (20 mA).
Help command: C1CP2=?
min <> max
if accepted
1:CMD ERR if current output not
enabled
5:ACCESS ERR if insufficient access
level
Read command: A1CP1?
Value of the cal. point if accepted
1:CMD ERR if analog input not
enabled
5:ACCESS ERR if insufficient access
level
A1CP1 (Analog input 1 Calibration
Point 1) Reads or sets the value of
the current output calibration point
1 (4 mA or min. voltage value).
Help command: A1CP1=?
min <> max
if accepted
1:CMD ERR if analog input not
enabled
5:ACCESS ERR if insufficient access
level
A1CP2 (Analog input 1 Calibration
Point 2) Reads or sets the value of
the current output calibration point
2 (20 mA or max. voltage value).
Read command: A1CP2?
Value of the cal. point if accepted
1:CMD ERR if analog input not
enabled
5:ACCESS ERR if insufficient access
level