Burkert Type 8056 User Manual
Page 30
- 30 -
RS232_485_ETP_MODBUS_BU_REV02.doc
3.3.14.
LIST OF SUPPORTED COMMANDS GROUPED SIMILARLY TO THE INTERNAL MENU
SYSTEM
NOTE ( VALID FOR ALL FUNCTIONS ) : returned value 5 = ACCESS ERR ( insufficient access
level )
COMMANDS AND FUNCTIONS EQUIVALENT TO THE “1-SENSOR” MENU
Name and description
Modes
Returned values or codes
Read command: PDIMV?
value of nominal diameter if accepted
5:ACCESS ERR if insufficient access
level
Set command: PDIMV=
n
0:OK if parameters accepted
2:PARAM ERR if
n
out of range
5:ACCESS ERR if insufficient access
level
PDIMV (Pipe DIaMeter Value)
Reads or sets the nominal diameter
of the sensor.
Help command: PDIMV=?
min <> max (unit)
if accepted
5:ACCESS ERR if insufficient access
level
Read command: CFFKA?
Value of the coefficient if accepted
5:ACCESS ERR if insufficient access
level
Set command: CFFKA=
n
0:OK if parameters accepted
2:PARAM ERR if
n
out of range
5:ACCESS ERR if insufficient access
level
CFFKA (CoeFFicient KA) Reads or
sets the value of the gain coefficient
KA
Help command: CFFKA=?
min <> max
if accepted
5:ACCESS ERR if insufficient access
level
Read command: SMODL?
Value of the sensor model if accepted
5:ACCESS ERR if insufficient access
level
5:ACCESS ERR if insufficient access
level
Set command: SMODL=
n
0:OK if parameters accepted
2:PARAM ERR if
n
out of range
5:ACCESS ERR if insufficient access
level
SMODL (Sensor MODeL) Reads or
sets the value of the sensor model
Help command: SMODL=?
min <> max
if accepted
5:ACCESS ERR if insufficient access
level
Read command: SIPOS?
Value of the insertion position if
accepted
5:ACCESS ERR if insufficient access
level
Set command: SIPOS=
n
0:OK if parameters accepted
2:PARAM ERR if
n
out of range
5:ACCESS ERR if insufficient access
level
SIPOS (Sensor Insertion POSition)
Reads or sets the value that
identifies the insertion position for
that type of sensor
Help command: SIPOS=?
min <> max
if accepted
5:ACCESS ERR if insufficient access
level
CKLP0, CKLP1, CKLP2, CKLP3,
CKLP4, CKLP5 (Coefficient KL
Positive 0, 1, 2, 3, 4 an 5) Reads
Read command: CKLP0?
Value of the coefficient if accepted
5:ACCESS ERR if insufficient access
level